Performance of Platelet Flags on the Sysmex XN-1000 Hematology Analyzer (Burkina Faso)
DOI:
https://doi.org/10.64294/jsd.v4i3.427Keywords:
Platelet, Blood Cell Count, Pseudothrombocytopenia, Sysmex XN-1000, Burkina FasoAbstract
Background: The complete blood count, performed by hematology analyzers, provides, in addition to numerical values, flags including those related to the platelet lineage. This study aimed to evaluate the performance of the platelet flags of the Sysmex XN-1000 analyzer in Burkina Faso.
Methods: This was a descriptive and evaluative cross-sectional study conducted over two months (November-December 2024) among patients referred for a complete blood count at the hematology laboratory of the Charles de Gaulle Pediatric University Hospital. Blood samples collected in EDTA tubes were analyzed on the Sysmex XN-1000 analyzer, and two blood smears were then prepared per patient for microscopic reading by two biologist hematologists, searching for abnormalities likely to interfere with platelet counting.
Results: A total of 389 samples were collected. The sensitivity, specificity, positive predictive value, negative predictive value, and total accuracy of the "PLT Clumps?" flag were 46.6% (95% CI: 38.7-54.6), 90.8% (95% CI: 86.3-94.2), 78.1% (95% CI: 68.5-85.9), 70.6% (95% CI: 65.1-75.8), and 72.5% (95% CI: 67.8-76.9), respectively. For the "Giant Platelet?" flag, these same parameters were 57.1% (95% CI: 34.0-78.2), 97.5% (95% CI: 95.4-98.9), 57.1% (95% CI: 34.0-78.2), 97.5% (95% CI: 95.4-98.9), and 95.4% (95% CI: 92.8-97.2), respectively. One case of platelet satellitism (0.3%) was also observed.
Conclusion: These results indicate that particular attention to these platelet alarms would be important during the biological validation of complete blood counts.